Publication
Title
4-French-compatible endovascular material is safe and effective in the treatment of femoropopliteal occlusive disease : results of the 4-EVER Trial
Author
Abstract
Purpose: To report the 1-year results of a prospective multicenter trial to evaluate the safety and efficacy of treating symptomatic femoropopliteal occlusive disease using 4-F-compatible materials and no closure device. Methods: The non-randomized 4-EVER trial (4-F endovascular treatment approach to infrainguinal disease) was conducted at 5 European hospitals (ClinicalTrials.gov identifier NCT01413139). The protocol mandated the use of only 4-F sheaths, self-expanding nitinol stents (Astron Pulsar or Pulsar-18 stent), and balloons from a single manufacturer. Between June 2010 and June 2011, 120 symptomatic patients (82 men; mean age 71 +/- 9.7 years, range 47-90), primarily claudicants, treated for 120 femoropopliteal lesions (>90% TASC A/B) were enrolled. The mean lesion length was 71.0 +/- 45.9 mm. Follow-up evaluations were scheduled on day 1 and at 1, 6, 12, and 24 months. A duplex ultrasound was performed on all follow-up visits to determine vessel patency (primary outcome measure at 1 year), and biplanar radiography was performed at 12 and 24 months to assess stent fracture. Results: Stents were successfully implanted in all patients: an Astron Pulsar stent in 70 (58.3%) lesions and a Pulsar-18 stent in 46(38.3%); 4(3.3%) patients had both stents implanted for flow-limiting dissection after predilation. No closure devices were used; the mean manual compression time was 8.1 minutes (2-15). Four (3.3%) patients developed significant hematoma at the puncture site, but none required surgical repair. The overall 12-month primary patency rate was 81.4%: 85.2% for the Astron Pulsar and 73.4% for the Pulsar-18 (p=0.236). Freedom from target lesion revascularization at 12 months for the entire cohort was 89.3%. Conclusion: Compared to published historical data for superficial femoral artery type A/B lesion stenting using 6-F devices, the 4-F devices applied in this trial showed similar patency at 12 months, fewer access site complications, and shorter manual compression times, supporting the supposition that 4-F endovascular treatment is safe and effective.
Language
English
Source (journal)
Journal of endovascular therapy
Publication
2013
ISSN
1526-6028
DOI
10.1583/13-4437MR.1
Volume/pages
20 :6 (2013) , p. 746-756
ISI
000328176500006
Full text (Publisher's DOI)
UAntwerpen
Faculty/Department
Research group
Publication type
Subject
Affiliation
Publications with a UAntwerp address
External links
Web of Science
Record
Identifier
Creation 15.01.2014
Last edited 04.03.2024
To cite this reference