Publication
Title
Influence of spark generator parameters in the analysis of graphite-electrodes by spark source-mass spectrometry
Author
Abstract
Spark source mass spectrometric analyses of doped carbon samples were performed at different pulse lengths, pulse frequencies and breakdown voltages. The pulse length and pulse frequencies had no influence on the analysis. The relation between breakdown voltage, spark gap and sample matrix was determined. With increasing breakdown voltage, the abundance of the multiply charged ions was found to decrease, whereas the abundance of the cluster ions increased. The breakdown voltage was also found to affect the absolute and relative intensities of different elements in different manners. Interpretation of these results led to some new considerations on the phenomena in a spark discharge.
Language
English
Source (journal)
International journal of mass spectrometry and ion processes. - Amsterdam, 1983 - 1998
Publication
Amsterdam : 1984
ISSN
0168-1176 [print]
1873-2801 [online]
DOI
10.1016/0168-1176(84)87087-1
Volume/pages
55 :3 (1984) , p. 233-247
ISI
A1984SA13700002