Publication
Title
Automated evaluation of photographically recorded spark-source mass spectra
Author
Abstract
A computer routine was developed for qualitative and quantitative analysis of photographically recorded spark-source mass spectra. Particular attention is given to the case of a graphite matrix. The program starts from the line intensities (expressed as Seidel values) and isotope masses calculated from the densitometer readings by a commercially available routine. From the intensities in the different exposures (typically 15 stages), it computes the parameters for the linear parts of the density curves for each ion. Taking into account mutual interferences of multivalent ions, isotope or C-clusters, oxide, carbide and dicarbide ions, the program automatically identifies and then quantifies the elements present. The precision of the results is around 5%. Reading and complete processing of one photoplate is achieved within 23 h.
Language
English
Source (journal)
Analytica chimica acta. - Amsterdam, 1947, currens
Publication
Amsterdam : 1978
ISSN
0003-2670 [print]
1873-4324 [online]
DOI
10.1016/S0003-2670(01)84041-1
Volume/pages
103 :3 (1978) , p. 223-35
ISI
A1978GD90900003