Title
|
|
|
|
Who is reviewing the data review systems of automated external defibrillators? Implications of flawed time lines for clinicians and researchers
|
|
Author
|
|
|
|
|
|
Abstract
|
|
|
|
The unintentional print-out by two different software programmes of the same resuscitation-related events stored on a data card of a Laerdal® FR2-automated external defibrillator (AED), led to the discovery of flaws in the registration of the time line by one of the commercially available Laerdal® software programmes. This observation stresses the need for a continuation of the medical supervision of AED projects, the close co-operation between clinicians and AED manufacturers, the well-controlled introduction of new devices and strict postmarket surveillance programmes. |
|
|
Language
|
|
|
|
English
|
|
Source (journal)
|
|
|
|
Resuscitation. - Limerick, 1972, currens
|
|
Publication
|
|
|
|
Limerick
:
2007
|
|
ISSN
|
|
|
|
0300-9572
[print]
1873-1570
[online]
|
|
DOI
|
|
|
|
10.1016/J.RESUSCITATION.2006.07.003
|
|
Volume/pages
|
|
|
|
72
:3
(2007)
, p. 484-489
|
|
ISI
|
|
|
|
000245069000020
|
|
Full text (Publisher's DOI)
|
|
|
|
|
|