Title
Who is reviewing the data review systems of automated external defibrillators? Implications of flawed time lines for clinicians and researchersWho is reviewing the data review systems of automated external defibrillators? Implications of flawed time lines for clinicians and researchers
Author
Faculty/Department
Faculty of Medicine and Health Sciences
Publication type
article
Publication
Limerick,
Subject
Human medicine
Source (journal)
Resuscitation. - Limerick, 1972, currens
Volume/pages
72(2007):3, p. 484-489
ISSN
0300-9572
1873-1570
ISI
000245069000020
Carrier
E
Target language
English (eng)
Full text (Publishers DOI)
Abstract
The unintentional print-out by two different software programmes of the same resuscitation-related events stored on a data card of a Laerdal® FR2-automated external defibrillator (AED), led to the discovery of flaws in the registration of the time line by one of the commercially available Laerdal® software programmes. This observation stresses the need for a continuation of the medical supervision of AED projects, the close co-operation between clinicians and AED manufacturers, the well-controlled introduction of new devices and strict postmarket surveillance programmes.
E-info
https://repository.uantwerpen.be/docman/iruaauth/95ed7f/0f38054e8cd.pdf
http://gateway.webofknowledge.com/gateway/Gateway.cgi?GWVersion=2&SrcApp=PARTNER_APP&SrcAuth=LinksAMR&KeyUT=WOS:000245069000020&DestLinkType=RelatedRecords&DestApp=ALL_WOS&UsrCustomerID=ef845e08c439e550330acc77c7d2d848
http://gateway.webofknowledge.com/gateway/Gateway.cgi?GWVersion=2&SrcApp=PARTNER_APP&SrcAuth=LinksAMR&KeyUT=WOS:000245069000020&DestLinkType=FullRecord&DestApp=ALL_WOS&UsrCustomerID=ef845e08c439e550330acc77c7d2d848
http://gateway.webofknowledge.com/gateway/Gateway.cgi?GWVersion=2&SrcApp=PARTNER_APP&SrcAuth=LinksAMR&KeyUT=WOS:000245069000020&DestLinkType=CitingArticles&DestApp=ALL_WOS&UsrCustomerID=ef845e08c439e550330acc77c7d2d848