Title
Who is reviewing the data review systems of automated external defibrillators? Implications of flawed time lines for clinicians and researchers
Author
Faculty/Department
Faculty of Medicine and Health Sciences
Publication type
article
Publication
Limerick ,
Subject
Human medicine
Source (journal)
Resuscitation. - Limerick, 1972, currens
Volume/pages
72(2007) :3 , p. 484-489
ISSN
0300-9572
1873-1570
ISI
000245069000020
Carrier
E
Target language
English (eng)
Full text (Publishers DOI)
Abstract
The unintentional print-out by two different software programmes of the same resuscitation-related events stored on a data card of a Laerdal® FR2-automated external defibrillator (AED), led to the discovery of flaws in the registration of the time line by one of the commercially available Laerdal® software programmes. This observation stresses the need for a continuation of the medical supervision of AED projects, the close co-operation between clinicians and AED manufacturers, the well-controlled introduction of new devices and strict postmarket surveillance programmes.
E-info
https://repository.uantwerpen.be/docman/iruaauth/95ed7f/0f38054e8cd.pdf
http://gateway.webofknowledge.com/gateway/Gateway.cgi?GWVersion=2&SrcApp=PARTNER_APP&SrcAuth=LinksAMR&KeyUT=WOS:000245069000020&DestLinkType=RelatedRecords&DestApp=ALL_WOS&UsrCustomerID=ef845e08c439e550330acc77c7d2d848
http://gateway.webofknowledge.com/gateway/Gateway.cgi?GWVersion=2&SrcApp=PARTNER_APP&SrcAuth=LinksAMR&KeyUT=WOS:000245069000020&DestLinkType=FullRecord&DestApp=ALL_WOS&UsrCustomerID=ef845e08c439e550330acc77c7d2d848
http://gateway.webofknowledge.com/gateway/Gateway.cgi?GWVersion=2&SrcApp=PARTNER_APP&SrcAuth=LinksAMR&KeyUT=WOS:000245069000020&DestLinkType=CitingArticles&DestApp=ALL_WOS&UsrCustomerID=ef845e08c439e550330acc77c7d2d848