Publication
Title
Heterogeneity effects in direct XRF analysis of traces of heavy metals preconcentrated on polyurethane foam sorbents
Author
Abstract
Trace amounts of heavy metals were preconcentrated on discs of polyurethane (PU) foam loaded with ammonium diethyldithiocarbamate (DDTC), to increase the sensitivity of XRF. Since the diameter of the cavities in the open-cell PU foams reaches 30300 μm, considerable heterogeneity effects might occur, which could reduce the accuracy of the method. The possible systematic error introduced by considering the sample to be homogeneous is calculated using the Berry et al. model. The calculations show that the underestimation of the absorption correction factor when considering the sample as being homogeneous is less than 2% for high-Z elements when the sample thickness is at least 2 mm. The detection limits are also around the minima for this thickness.
Language
English
Source (journal)
X-ray spectrometry. - London
Publication
London : 1986
ISSN
0049-8246
DOI
10.1002/XRS.1300150104
Volume/pages
15 :1 (1986) , p. 7-11