Title
Impact of spatial resolution on results of esophageal high-resolution manometryImpact of spatial resolution on results of esophageal high-resolution manometry
Author
Faculty/Department
Faculty of Medicine and Health Sciences
Research group
Gastroenterologie
Publication type
article
Publication
Cambridge, Mass.,
Subject
Human medicine
Source (journal)
Neurogastroenterology and motility / European Gastrointestinal Motility Society. - Cambridge, Mass., 1994, currens
Volume/pages
26(2014):7, p. 922-928
ISSN
1350-1925
ISI
000338024600004
Carrier
E
Target language
English (eng)
Full text (Publishers DOI)
Affiliation
University of Antwerp
Abstract
Background The Chicago classification for esophageal motility disorders was designed for a 36-channel manometry system with sensors spaced at 1cm. However, many motility laboratories outside the USA use catheters with a lower resolution in the segments outside the esophagogastric junction. Our aim was to investigate the effect of spatial resolution on the Chicago metrics and diagnosis. Methods In 20 healthy volunteers and 47 patients with upper gastrointestinal symptoms, high-resolution manometric studies of the esophagus were retrospectively reanalyzed using the original 1-cm spacing in the segments outside the 7-cm esophagogastric junction segment, and again after manually increasing the spacing between sensors to 2, 3, and 4cm (above the lower esophageal sphincter region). Measurements were analyzed in random order and the investigator was blinded to the outcome of the analyses performed in another resolution of the same patient. Intra-class correlation coefficients (ICC) and Kappa values were determined. Key Results There was a very strong correlation between the 1-cm and 2-cm analysis for all Chicago metrics studied in healthy volunteers (ICCs: distal contractile integral 0.998; contractile front velocity (CFV) 0.964; distal latency 0.919; peristaltic break size 0.941). The 2-cm spacing analysis also correlated very well with the 1-cm analysis for the different Chicago diagnoses obtained in the patients (Kappa values ranging from 0.665 to 1.000). When the sensor spacing was increased to 3 and 4cm, the correlation was reduced to moderate for the Chicago metrics, especially for break size and CFV of peristalsis. Conclusions & Inferences The Chicago classification for esophageal motility disorders is still valid and the same normal values can be used when catheters with a slightly lower resolution are used (i.e., 2-cm vs 1-cm spacing). For larger sensor intervals, the classification and the normal values will need to be adjusted.
E-info
https://repository.uantwerpen.be/docman/iruaauth/ba0ab5/7477897.pdf
http://gateway.webofknowledge.com/gateway/Gateway.cgi?GWVersion=2&SrcApp=PARTNER_APP&SrcAuth=LinksAMR&KeyUT=WOS:000338024600004&DestLinkType=RelatedRecords&DestApp=ALL_WOS&UsrCustomerID=ef845e08c439e550330acc77c7d2d848
http://gateway.webofknowledge.com/gateway/Gateway.cgi?GWVersion=2&SrcApp=PARTNER_APP&SrcAuth=LinksAMR&KeyUT=WOS:000338024600004&DestLinkType=FullRecord&DestApp=ALL_WOS&UsrCustomerID=ef845e08c439e550330acc77c7d2d848
http://gateway.webofknowledge.com/gateway/Gateway.cgi?GWVersion=2&SrcApp=PARTNER_APP&SrcAuth=LinksAMR&KeyUT=WOS:000338024600004&DestLinkType=CitingArticles&DestApp=ALL_WOS&UsrCustomerID=ef845e08c439e550330acc77c7d2d848
Handle