Hardware prominence after plate fixation for clavicle fracture is a common complication. The aim of the study was to perform a 3D analysis of the prominence of different types of superior clavicle plates. An automated fitting of 3 straight and 10 precontoured plates was performed on 52 3D-CT-scan reconstructed cadaver clavicles. The mean and maximum bone-plate distance and maximum prominence was significant higher with the straight plates compared to the precontoured plates. The mean and maximum boneplate distance was significant higher with the precontoured DePuy-Synthes plates compared to the precontoured Acumed plates but when evaluating the maximum prominence there was no significant difference between the most commonly used 8-holes plates. To conclude, precontoured plates of the clavicula diminish significantly hardware prominence. There exists a difference in hardware prominence between different brands of precontoured plates but this difference is limited and in most cases not significant. |
| |