Abstract
Purpose: To evaluate the short- and midterm outcomes of the Viabahn endoprosthesis with Propaten Bioactive Surface vs. standard balloon angioplasty for treatment of in-stent restenosis in the superficial femoral artery. Methods: Between June 2010 and February 2012, 83 patients with superficial femoral artery in-stent and Rutherford category 2 to 5 ischemia were enrolled at 7 sites participating in this prospective randomized controlled study (RELINE; ClinicalTrials.gov; identifier NCT01108861). The patients were randomized to treatment with either the heparin-bonded Viabahn endoprosthesis (n=39: 29 men; mean age 67.7 +/- 9.8 years) or a standard angioplasty balloon (n=44: 32 men; mean age 69.0 +/- 9.7 years). The primary effectiveness outcome was primary patency at 12 months, defined as no restenosis/occlusion within the target lesion based on duplex ultrasound and no target lesion revascularization. The primary safety endpoint was the incidence of serious device-related adverse events within 30 days of the procedure. Results: All patients randomized to receive dilation or the Viabahn stent-graft were treated according to their assignment. The technical success was 100% for the Viabahn group and 81.8% for the angioplasty group (p=0.002) owing to 9 patients requiring a bailout procedure after unsuccessful angioplasty. The 12-month primary patency rates were 74.8% for the Viabahn group and 28.0% for the angioplasty group (p<0.001). Excluding the 9 angioplasty patients who received bailout stenting, the primary patency for optimal balloon angioplasty was 37.0% (p<0.001). Three patients experienced device-related adverse events within 30 days: occlusion of the target lesion (Viabahn group), peripheral embolization (angioplasty group), and reocclusion of the target lesion (angioplasty group). Conclusion: In this study, the treatment of femoropopliteal in-stent restenosis with a Viabahn endoprosthesis showed significantly better results than treatment with a standard balloon at 1 year.
Language
English
Source (journal)
Journal of endovascular therapy
Publication
2015
ISSN
1526-6028
DOI
10.1177/1526602814564385
Volume/pages
22 :1 (2015) , p. 1-10
ISI
000351413800001
Pubmed ID
25775672
Full text (Publisher's DOI)
UAntwerpen
Faculty/Department
Research group
Publication type
Subject
Affiliation
Publications with a UAntwerp address
External links
Web of Science
Record
Identifier
Creation 12.05.2015
Last edited 04.03.2024
To cite this reference