Title
An accurate projection model for diffraction image formation and inversion using a polychromatic cone beam An accurate projection model for diffraction image formation and inversion using a polychromatic cone beam
Author
Faculty/Department
Faculty of Sciences. Physics
Publication type
article
Publication
Copenhagen ,
Subject
Chemistry
Source (journal)
Journal of applied crystallography / International Union of Crystallography. - Copenhagen
Volume/pages
48(2015) :2 , p. 334-343
ISSN
0021-8898
ISI
000352229100003
Carrier
E
Target language
English (eng)
Full text (Publishers DOI)
Affiliation
University of Antwerp
Abstract
Recently, the concept of X-ray diffraction contrast tomography (DCT) has been extended to the case of more widely available laboratory source CT systems. Using well known concepts from geometrical ray optics, an exact formulation is derived for the forward and backward projection geometry encountered under polychromatic cone beam illumination, and it is shown how this projection model can be efficiently implemented in practice. The new projection model is subsequently used for iterative tomographic reconstruction of the three-dimensional shape of a grain from a set of experimentally observed cone beam projections and shows a clear improvement compared to the simplified projection model used previously.
E-info
http://gateway.webofknowledge.com/gateway/Gateway.cgi?GWVersion=2&SrcApp=PARTNER_APP&SrcAuth=LinksAMR&KeyUT=WOS:000352229100003&DestLinkType=RelatedRecords&DestApp=ALL_WOS&UsrCustomerID=ef845e08c439e550330acc77c7d2d848
http://gateway.webofknowledge.com/gateway/Gateway.cgi?GWVersion=2&SrcApp=PARTNER_APP&SrcAuth=LinksAMR&KeyUT=WOS:000352229100003&DestLinkType=FullRecord&DestApp=ALL_WOS&UsrCustomerID=ef845e08c439e550330acc77c7d2d848
http://gateway.webofknowledge.com/gateway/Gateway.cgi?GWVersion=2&SrcApp=PARTNER_APP&SrcAuth=LinksAMR&KeyUT=WOS:000352229100003&DestLinkType=CitingArticles&DestApp=ALL_WOS&UsrCustomerID=ef845e08c439e550330acc77c7d2d848
Handle