Title
Superior efficacy of pulmonary vein isolation with online contact force measurement persists after the learning period : a prospective case control study Superior efficacy of pulmonary vein isolation with online contact force measurement persists after the learning period : a prospective case control study
Author
Faculty/Department
Faculty of Medicine and Health Sciences
Faculty of Pharmaceutical, Biomedical and Veterinary Sciences . Biomedical Sciences
Publication type
article
Publication
,
Subject
Human medicine
Source (journal)
Journal of interventional cardiac electrophysiology. - Place of publication unknown
Volume/pages
43(2015) :3 , p. 287-296
ISSN
1383-875X
ISI
000358652300010
Carrier
E
Target language
English (eng)
Full text (Publishers DOI)
Affiliation
University of Antwerp
Abstract
Use of online contact force (CF) measurement during circumferential pulmonary vein (PV) isolation (CPVI) for atrial fibrillation (AF) has demonstrated improvements in procedural parameters and mid-term clinical outcome. However, it is unknown if experience gained with CF measuring catheters improves the efficacy of subsequent CPVI procedures performed without CF measurement. This prospective trial compared procedural results of CPVI performed without a CF measuring catheter to a control group performed with a CF measuring catheter, by an operator with prior experience with CF technology.. Thirty-six eligible paroxysmal (n = 27) or persistent (n = 9) AF patients were consecutively enrolled. Twelve patients underwent CPVI with the non-CF catheter (CF- group) in a recall period and 24 with the CF catheter (CF+ group). After the first circumferential lesion set, the number of PV pairs requiring additional touch-up lesions to achieve electrical isolation was significantly less in the CF+ group (2 of 48 (4.2 %) vs. 7 of 24 (29.2 %) in the CF+ and CF- groups, respectively, p = 0.005). The procedure time was significantly lower in the CF+ group (117.9 +/- 23.3 vs. 134.1 +/- 25.3 min, p = 0.033). Radiofrequency (RF) and fluoroscopy time did not differ between groups (31.5 +/- 7.1 vs. 31.8 +/- 7.0 min and 11.8 +/- 5.6 vs. 11.0 +/- 5.8 min in the CF+ and the CF- group, respectively) With the use of online CF measurement, PV isolation is more frequently complete following the first circumferential lesion set. A previous learning period with direct CF feedback is not a substitute for real-time direct CF measurement to maintain this advantage.
E-info
http://gateway.webofknowledge.com/gateway/Gateway.cgi?GWVersion=2&SrcApp=PARTNER_APP&SrcAuth=LinksAMR&KeyUT=WOS:000358652300010&DestLinkType=RelatedRecords&DestApp=ALL_WOS&UsrCustomerID=ef845e08c439e550330acc77c7d2d848
http://gateway.webofknowledge.com/gateway/Gateway.cgi?GWVersion=2&SrcApp=PARTNER_APP&SrcAuth=LinksAMR&KeyUT=WOS:000358652300010&DestLinkType=FullRecord&DestApp=ALL_WOS&UsrCustomerID=ef845e08c439e550330acc77c7d2d848
https://repository.uantwerpen.be/docman/iruaauth/d21053/291127174.pdf
Handle