Publication
Title
Epicardial pacing : a single-centre study on 321 leads in 138 patients
Author
Abstract
Objective - This study presents the long-term outcome of 321 epicardial leads in 138 patients. Methods and results - All leads were Medtronic CapsureEpi model 4965 steroid eluting leads. The 1-, 3-, and 5-year patient survival was 91%,83% and 77%, respectively. Twenty-seven patients died. In 25/27 deaths a pacing-related death could be excluded. Strangulation of the heart by an abandoned epicardial lead was the cause of death in one child. One other patient died suddenly at the age of 3 years. Failures occurred in 57 of 321 epicardial leads (18%). For all 321 leads, the 1-, 3- and 5-year freedom from failure was 91%, 85% and 71%, respectively. The cumulative proportion of patients without any lead defect was 85% after 1 year, 76% after 3 years and 62% after 5 years. The percentage of patients without serious adverse events at 1, 3, and 5 years was 97%, 91%, 85%, respectively. Lead fracture was the cause of failure in 15 leads of 9 patients. An important increase in pacing threshold occurred in 35 leads of 30 patients. Other failures were: diaphragmatic stimulation, infection, excessive traction and strangulation. Eighteen failures were repaired by 11 surgical interventions in 9 patients. Thirty-nine defects were corrected non-invasively in 31 patients. Conclusions - The use of steroid-eluting epicardial leads has proven to be an adequate option. In paediatric cardiology, the epicardial approach remains an indispensable tool for achieving a lifelong pacing.
Language
English
Source (journal)
Acta cardiologica. - Bruxelles
Publication
Bruxelles : 2006
ISSN
0001-5385
DOI
10.2143/AC.61.3.2014839
Volume/pages
61 :3 (2006) , p. 343-351
ISI
000238682200015
Full text (Publisher's DOI)
Full text (open access)
UAntwerpen
Publication type
Subject
External links
Web of Science
Record
Identifier
Creation 28.11.2016
Last edited 22.02.2023
To cite this reference