Peripheral doses in radiotherapy: A comparison between IMRT, VMAT and Tomotherapy
Author
Abstract
The goal of this intercomparison is to determine the peripheral doses during treatment of prostate and head and neck (H&N) cancers. In the case of prostate cancer, two different treatment techniques are compared: intensity-modulated radiation therapy (IMRT 10 MV and 18 MV), on a Varian Clinac 2100 C/ D and Tomotherapy. VMAT (also on a Varian Clinac 2100 C/D) was compared to Tomotherapy, for H&N cancer. The treatment devices are located at the university hospitals of Leuven and Brussels, respectively. A common treatment protocol was agreed between the two clinical centers and this same protocol was used by each partner. For the higher energy modalities (10 MV and 18 MV) we also assessed the neutron contribution to the total dose, by using bubble detectors. In this way, the performance (in terms of peripheral doses) of the different treatment techniques, when faced with the same dose distribution constraints, was evaluated. The doses were evaluated with an anthropomorphic phantom loaded with TLD detectors. Summarizing our results, we can conclude that low energy radiation techniques, namely VMAT and Tomotherapy, have more interesting performances when compared to IMRT at energies of 10 MV and 18 MV, with respect to peripheral dose. On the one hand the former are associated with lower photon doses and, on the other hand, there is no contribution from neutrons to the total dose. (C) 2013 Elsevier Ltd. All rights reserved.
Language
English
Source (journal)
Radiation measurements. - Oxford, 1994, currens
Source (book)
Workshop on Dosimetry for Second Cancer Risk Estimation at the EURADOS, Annual Meeting, FEB 08, 2012, IAEA, IAEA, Vienna, AUSTRIA
Publication
Oxford : Pergamon-elsevier science ltd , 2013
ISSN
1350-4487
DOI
10.1016/J.RADMEAS.2013.04.016
Volume/pages
57 (2013) , p. 62-67
ISI
000327284800008
Full text (Publisher's DOI)
UAntwerpen
Publication type
Subject
External links
Web of Science
Record
Identifier
Creation 04.12.2019
Last edited 20.08.2024
To cite this reference