Publication
Title
Setup accuracy of stereoscopic X-ray positioning with automated correction for rotational errors in patients treated with conformal arc radiotherapy for prostate cancer
Author
Abstract
We evaluated setup accuracy of NovalisBody stereoscopic X-ray positioning with automated correction for rotational errors with the Robotics Tilt Module in patients treated with conformal arc radiotherapy for prostate cancer. The correction of rotational errors was shown to reduce random and systematic errors in all directions. (NovalisBody (TM) and Robotics Tilt Module (TM) are products of BrainLAB A.G., Heimstetten, Germany). (c) 2006 Elsevier Ireland Ltd. All rights reserved.
Language
English
Source (journal)
Radiotherapy and oncology. - Amsterdam
Publication
Amsterdam : 2006
ISSN
0167-8140
DOI
10.1016/J.RADONC.2006.07.001
Volume/pages
80 :3 (2006) , p. 371-373