Publication
Title
A simple theoretical verification of monitor unit calculation for intensity modulated beams using dynamic mini-multileaf collimation
Author
Abstract
A spreadsheet based program is presented to perform an independent Monitor Unit (MU) calculation verification for the Quality Assurance (QA) of Intensity Modulated Radiation Therapy (IMRT) using Dynamic MultiLeaf Collimation (DMLC). The computed dose value is compared to the planned dose by calculating the percent dose difference per Intensity Modulated Beam (IMB) and absolute dose difference per IMB. The proposed acceptability levels are +/-1-5.0% or +/-2.0cGy for the percent dose difference per IMB and the absolute dose difference per IMB, respectively. For percent dose difference per treatment, an acceptability level of +/-2.0% is proposed. The presented program is considered adequate for checking the treatment plans calculated for IMRT treatments using DMLC as a part of the QA procedure. (C) 2004 Elsevier Ireland Ltd. All rights reserved.
Language
English
Source (journal)
Radiotherapy and oncology. - Amsterdam
Publication
Amsterdam : 2004
ISSN
0167-8140
DOI
10.1016/J.RADONC.2004.02.014
Volume/pages
71 :2 (2004) , p. 235-241
ISI
000221371300015