Title
|
|
|
|
Microprocessor controlled limitation system for a stand-alone freely movable treatment couch
| |
Author
|
|
|
|
| |
Abstract
|
|
|
|
Because of the capability of free movement in the treatment room, we recently introduced a Hercules treatment couch on one of our linear accelerators. One of the advantages of this couch is that it allows for a more flexible way of patient setup and that it can be moved entirely out of the way to enable treatment with a hospital bed. A disadvantage, however, is that the couch can hit a wall or a cover of the accelerator accidentally. A limitation system has been developed to protect both the table and the accelerator against such collisions. (C) 1998 American Association of Physicists in Medicine. [S0094-2405(98)02106-3]. |
| |
Language
|
|
|
|
English
| |
Source (journal)
|
|
|
|
Medical physics. - New York, N.Y.
| |
Publication
|
|
|
|
New York, N.Y.
:
1998
| |
ISSN
|
|
|
|
0094-2405
[print]
2473-4209
[online]
| |
DOI
|
|
|
|
10.1118/1.598267
| |
Volume/pages
|
|
|
|
25
:6
(1998)
, p. 897-899
| |
ISI
|
|
|
|
000074282600015
| |
Full text (Publisher's DOI)
|
|
|
|
| |
|