Publication
Title
Probe integrated scattering cross sections in the analysis of atomic resolution HAADF STEM images
Author
Abstract
The physical basis for using a probe-position integrated cross section (PICS) for a single column of atoms as an effective way to compare simulation and experiment in high-angle annular dark-field (HAADF) scanning transmission electron microscopy (STEM) is described, and the use of PICS in order to make quantitative use of image intensities is evaluated. It is based upon the calibration of the detector and the measurement of scattered intensities. Due to the predominantly incoherent nature of HAADF STEM, it is found to be robust to parameters that affect probe size and shape such as defocus and source coherence. The main imaging parameter dependencies are on detector angle and accelerating voltage, which are well known. The robustness to variation in other parameters allows for a quantitative comparison of experimental data and simulation without the need to lit parameters. By demonstrating the application of the [PICS to the chemical identification of single atoms in a heterogeneous catalyst and in thin, layered-materials, we explore some of the experimental considerations when using this approach. (C) 2013 Elsevier B.V. All rights reserved.
Language
English
Source (journal)
Ultramicroscopy. - Amsterdam
Publication
Amsterdam : 2013
ISSN
0304-3991
DOI
10.1016/J.ULTRAMIC.2013.07.002
Volume/pages
133 (2013) , p. 109-119
ISI
000324471800015
Full text (Publisher's DOI)
UAntwerpen
Publication type
Subject
External links
Web of Science
Record
Identifier
Creation 17.02.2020
Last edited 09.12.2024
To cite this reference