Publication
Title
A new technique for difficult removal of a peripherally inserted central venous catheter (PICC) in a neonate
Author
Abstract
In general, the removal of peripherally inserted central venous catheters (PICC) in neonates by gentle traction is easy. In our case, the removal of a 28G PICC in a term neonate was impossible by manual traction even with force. Previously described non-invasive interventions using a stylet were not successful because it was not possible to pass the stylet along the catheter hub of the narrow 28G PICC. In the end, the catheter could be removed non-operatively by cutting the catheter just distal to the hub and inserting a stylet of a new PICC (same brand and size) into the patients' retained catheter. Subsequently, the force of manual traction on the catheter could be increased without increased risk of catheter stretching and breakage. After catheter removal, the surface of the remaining PICC was intact. Conclusion: By thinking outside the box, surgical intervention was prevented in this neonate.
Language
English
Source (journal)