Publication
Title
Interface for reproducible, multishot direct analysis of solid-phase microextraction samples
Author
Abstract
An enclosed interface that joins a direct analysis in real time (DART) probe, solid-phase microextraction (SPME) fiber, and the inlet of a high-resolution mass spectrometer is described. Unlike other systems to couple SPME sampling to ambient mass spectrometry, the interface is able to perform discrete analyses on different areas of a single SPME fiber device for up to three technical replicate measurements of one sampling event. Inlet flow speed and desorption temperature are optimized, and reproducibility is demonstrated between replicate analyses on the same derivatized SPME fiber and with sequential fiber sampling events, yielding analyte measurement center of variance (CV) from 3 to 6%. Conditioning is also performed with the enclosed DART. The interface is a straightforward addition to commercially available technologies, and machine diagrams for custom components operated with SPME/DART/MS equipment are included.
Language
English
Source (journal)
Analytical chemistry. - Washington, D.C., 1948, currens
Publication
Washington, D.C. : 2020
ISSN
0003-2700 [print]
5206-882X [online]
DOI
10.1021/ACS.ANALCHEM.9B05691
Volume/pages
92 :6 (2020) , p. 4182-4186
ISI
000526563900004
Pubmed ID
32030986
Full text (Publisher's DOI)
Full text (publisher's version - intranet only)
UAntwerpen
Research group
Publication type
Subject
External links
Web of Science
Record
Identifier
Creation 08.10.2021
Last edited 24.08.2024
To cite this reference