Title
|
|
|
|
Effect of changing pulse rate on profile parameters of perceptual thresholds and loudness comfort levels and relation to ECAP thresholds in recipients of the Nucleus CI24RE device
|
|
Author
|
|
|
|
|
|
Abstract
|
|
|
|
The Nucleus CI24RE Freedom device offers higher stimulation rates and lower noise levels in action potential measurements (ECAPs) than previous devices. A study including ten European implant teams showed that the effect of changes in rate from 250 to 3500 pulses per second on tilt and curvature of the T and C profiles is insignificant. When changing rate one may change the levels at all electrodes by the same amount. Using an automated procedure ECAPs could be measured quickly and reliably at a noise level of only 1 μV. However, this did not result in improved correlations between the tilt and curvature parameters of the ECAP profiles and those of the T and C profiles. Average C levels appear to differ markedly among implant centers; a better assessment protocol is required. When increasing stimulus rate one should take into account that this requires higher pulse charges per second and more power consumption. |
|
|
Language
|
|
|
|
English
|
|
Source (journal)
|
|
|
|
The international journal of audiology / Royal National Institute for the Deaf; British Society of Audiology. - Stockholm
|
|
Publication
|
|
|
|
Stockholm
:
2010
|
|
ISSN
|
|
|
|
1499-2027
[print]
1708-8186
[online]
|
|
DOI
|
|
|
|
10.3109/14992027.2010.492401
|
|
Volume/pages
|
|
|
|
49
:10
(2010)
, p. 775-787
|
|
ISI
|
|
|
|
000282817200008
|
|
Full text (Publisher's DOI)
|
|
|
|
|
|