Title
|
|
|
|
Design of a high resolution scintillator based SPECT detector (SPECTatress)
|
|
Author
|
|
|
|
|
|
Abstract
|
|
|
|
SPECT scanners using multi-pinhole collimators benefit from compact detectors having a high spatial resolution. Such detectors can be placed closer to the collimator and perpendicular to the pinhole axis (thereby limiting DOI spatial resolution degradation). Current clinical gamma ray cameras have a large area and a poor spatial resolution. This proceeding describes the architecture of SPECTatress, a compact high resolution gamma camera. The main components (PMT and scintillator) are presented, a brief overview of the read-out electronics is given and a first measurement using the center-of-gravity (COG) event-positioning algorithm is shown. The result of this measurement indicates a good spatial resolution (1.75 mm FWHM) in the center of the detector. Further research has to be done on the event-positioning algorithm to extend the usable detector area and to enhance the spatial resolution. |
|
|
Language
|
|
|
|
English
|
|
Source (journal)
|
|
|
|
Nuclear instruments and methods in physics research : A: accelerators, spectrometers, detectors and associated equipment. - Amsterdam, 1984, currens
|
|
Publication
|
|
|
|
Amsterdam
:
2011
|
|
ISSN
|
|
|
|
0168-9002
|
|
DOI
|
|
|
|
10.1016/J.NIMA.2010.12.081
|
|
Volume/pages
|
|
|
|
648
:S:1
(2011)
, p. S107-S110
|
|
ISI
|
|
|
|
000305376900029
|
|
Full text (Publisher's DOI)
|
|
|
|
|
|
Full text (publisher's version - intranet only)
|
|
|
|
|
|