Publication
Title
Simultaneous maximum a posteriori reconstruction of attenuation and activity distributions from emission sinograms
Author
Abstract
In order to perform attenuation correction in emission tomography an attenuation map is required. The authors propose a new method to compute this map directly from the emission sinogram, eliminating the transmission scan from the acquisition protocol. The problem is formulated as an optimization task where the objective function is a combination of the likelihood and an a priori probability. The latter uses a Gibbs prior distribution to encourage local smoothness and a multimodal distribution for the attenuation coefficients. Since the attenuation process is different in positron emission tomography (PET) and single photon emission tomography (SPECT), a separate algorithm for each case is derived. The method has been tested on mathematical phantoms and on a few clinical studies. For PET, good agreement was found between the images obtained with transmission measurements and those produced by the new algorithm in an abdominal study. For SPECT, promising simulation results have been obtained for nonhomogeneous attenuation due to the presence of the lungs.
Language
English
Source (journal)
IEEE transactions on medical imaging / Institute of Electrical and Electronics Engineers [New York, N.Y.] - New York, N.Y.
Publication
New York, N.Y. : 1999
ISSN
0278-0062
Volume/pages
18:5(1999), p. 393-403
ISI
000081371900004
Full text (Publishers DOI)